PRE-GELLED DISPOSABLE ELECTRODES FOR ELECTROCARDIOGRAPHY

The FIAB pre-gelled disposable electrode F3001ECG has technical characteristics that make it completely reliable during ECG electrocardiographic signal recordings. It is radiolucent and therefore allows for radiographic procedures without removing the electrode. None of the components contain any ferro-magnetic materials, therefore these electrodes are compatible with applications in MRI environments. The electrodes are tested for such compatibility with magnetic field levels up to 4.7 T. The use of electrodes does not affect the safety and operation of MRI systems. It is compatible with any electrocardiograph on the market, is pre-coated with solid gel and is equipped with a connection suitable for connection to cables with crocodile clips. Compliant with ANSI / AAMIEC12: 2000 standard, it allows the recovery of the ECG signal after discharge from the defibrillator.
The product does not contain latex.

TECHNICAL FEATURES

DESCRIPTION Disposable electrode pregellated with solid gel for ECG
DIMENSIONS 34 x 23 mm
GEL Solid state biocompatible medical adhesive gel
CONDUCTOR MATERIAL Ag / AgCl SSC
LINK Direct connection with crocodile
COMPLIANCE WITH REGULATIONS Directive 93/42 / EEC (Legislative Decree 46/97) and subsequent changes.
Class I medical device
ANSI / AAMI EC12: 2000 "Disposable ECG Electrodes"

ELECTRICAL FEATURES

ANSI / AAMIEC12: 2000 SPECIFICATION MAXIMUM VALUE
ANSI / AAMI
RESULT
OF TESTS
Average value of impedance measured at 10 Hz for 12 pairs of electrodes 2 kΩ 534 Ω
Impedance of the single pair at 10 Hz 3 kΩ 610 kΩ
Offset voltage 0.79 mV 100 mV
Combined offset instability and inherent noise 150 mcV 54 mcV
Residual voltage after defibrillation (5 sec) 19.8 mV 100 mV
Rate of polarization variation 1 mV / sec 0.21mV / sec
Impedance at 10 Hz after the test 3kΩ 197 kΩ
Tolerance on the polarization current 100 mV 19.1 mV

storage

Temperature: + 5 ° C ÷ + 30 ° C
Relative humidity: 20% ÷ 80%
EXPIRY: 36 months from the date of production

Format

Pack of 100 pieces.
Cod. F3001 / ECG