Shipping from as low as 8,00€

Customer care
  • VTRUST TD-4279 Glucose 25 Strips -49%

VTRUST TD-4279 Glucose 25 Strips

19.50€ 38.00€
  • Purchase 3 items for 19.11€ each
  • Purchase 4 items for 18.72€ each
  • Purchase 5 items for 18.33€ each
Price information ⓘ

VTRUST TD-4279 - Test Strips

Blood glucose test strips for in vitro diagnostic use. Single use.

VTRUST TD-4279 Test Strips, when used in conjunction with the VTRUST TD-4279 Beta Ketone and Glucose Monitoring System, allow for measurement of blood levels by a home user or healthcare professional. The system uses fresh finger whole blood samples. This system is not intended to be used for the diagnosis or screening of diabetes mellitus.
Professionals can use the test strips to test capillary, venous and arterial blood samples; home use is limited to the capillary whole blood test.
EN ISO 15197: 2015 indicates that the results provided by glucose meters fall within ± 15 mg / dL (0.83 mmol / L) of the value of the reference method when the glucose concentration is less than 100 mg / dL ( 5.55mmol / L), and are within ± 15% of the reference method value when the glucose concentration is 100 mg / dL (5.55 mmol / L) or higher. The following tables display how often the VTRUST TD-4279 achieves this goal. The table is based on a study carried out in 160 patients (each patient was tested six times thus obtaining 960 results) to evaluate the performance of the VTRUST TD-4279 instrument system and VTRUST TD-4279 (TD-4330) strips compared to the YSI reference method -2300.

Accuracy of results for glucose concentrations <100 mg / dL (5.55 mmol / L) Within ± 5mg / dL
(Within ± 0.27 mmol / L) Within ± 10 mg / dL
(Within ± 0.55 mmol / L) Within ± 15 mg / dL
(Within ± 0.83 mmol / L) 83.3% (270/324) 100% (324/324) 100% (324/324)
Accuracy of results for glucose concentrations ≥ 100 mg / dL (5.55 mmol / L)

Within ± 5% Within ± 10% Within ± 15%
66.4% (422/636) 96.9% (616/636) 100% (636/636)


Accuracy of results for glucose concentrations between 31.2 mg / dL (1.73 mmol / L) and 537 mg / dL (36.06 mmol / L)

Within ± 15 mg / dL or ± 15%
(Within ± 0.83 mmol / L or ± 15%)
100% (960/960)


The CV (%) is less than 5% for both intermediate precision and repeatability.

How to use

Wash and dry your hands before testing.
Insert the test strip all the way into the slot on the meter until it proceeds further. When the strip is fully inserted, the tool will perform several checks. Collect a blood sample of approximately 0.5 microliters with the test strip. Sufficient blood must be applied for accurate results. Bring the drop of blood into contact with the absorbent hole of the test strip and wait until the confirmation window is completely filled. Do not apply a finger blood sample by smearing it on the strip. The device will start the countdown. After a few seconds the instrument will display the blood glucose value. The last reading will be automatically saved in the instrument. Turn off the device by removing the test strip and throw away the test strip.
The test results may be incorrect if the contact bar is not fully inserted

Format

Pack of 25 strips.